DPN 402197 © TSS (International) Ltd Page 7 of 10
amp gain and autogain to default settings. Repeat steps 2 and 3 using the values
outlined below.
17.Press [9] to exit the Terminal Menu.
18.To re-establish communication with the 350 System, press the Enable/disable
polling button on the Terminal Window toolbar. Scrolling data will be displayed on
screen to identify the system is operating correctly.
E.2.1 Frequency Selection
Tests can be run at different frequencies ranging from 25Hz to 21Hz. To do this,
remove the Coil Tester end cap and adjust the Frequency Selection Switch.
When supplied by TSS, the Coil Tester will be set to its default position five and 25Hz.
It is strongly recommended that a default test is carried out using this frequency set-
ting.
It is important to be aware that changing the frequency will change the coil voltage
results achieved. The expected coil voltages for the available frequency outputs are
explained in Table E–4.
Table E–3: 350 System Operating Parameters
Parameter Required Setting
Pre-amp Gain 4
Pre-amp Autogain ON
Table E–4: Coil Tester Frequency settings and expected coil voltages
Switch Position Frequency Output Coil Voltage
5 25Hz +/- 0.0025Hz 1.0 to 1.5e6
4 24Hz +/- 0.0025Hz 0.94 to 1.44e6
3 23Hz +/- 0.0025Hz 0.89 to 1.39e6
2 22Hz +/- 0.0025Hz 0.83 to 1.33e6
1 21Hz +/- 0.0025Hz 0.79 to 1.29e6